Noncontact Mode

In noncontact mode, the long-distance van der Waals forces are sensed by an oscillating probe, which is excited at its resonance frequency by a piezoelectric transducer and brought in close proximity to a sample. The probe oscillation amplitudes are typically in the range of 1 nm or less. The resolution in this mode is defined by the tip-sample separation of a few nanometers. For the scanning feedback either the vibration amplitude (amplitude modulation mode, AM) or the frequency shift (frequency modulation, FM) of the probe can be used.

Fig.1. Microdrops of a liquid n-alkane observed by AFM in noncontact mode. Image courtesy of A. Checco.
In Fig. 1 an image of microdrops of a liquid n-alkane on a silanized silicon substrate is shown. The use of noncontact mode allowed scanning without influencing the shape of the microdrops by tip-sample forces.

To improve resolution in noncontact mode, enhancement of the sensitivity to short-range tip-sample forces is needed. To do this, small amplitudes of oscillation and small tip-sample distances are commonly used. Advances in the development of the noncontact mode allow for the discovery of single defects in atomic lattices. An image of the Si 7x7 surface obtained in noncontact mode is presented in Fig. 2.

 Fig.2. AFM image of Si (111) 7x7 acquired in noncontact mode on JEOL JSPM-4500A.

Operation at distances near the sample demands fine tip-force control and the absence of surface contamination. That is why this technique is mostly used in ultrahigh vacuum.

In most cases, the cantilever of choice for this mode is one with a high spring constant in the range 20 - 100 N/m so that it does not stick to the sample surface at small amplitudes. Between two cantilevers having the same spring constant, it is better to use the one with the higher resonant frequency. This lever is generally faster and less noisy. Low-frequency cantilevers are used with AFM systems that do not support probes with short lever arms.

Not surprisingly, imaging in noncontact mode is performed with Si probes usually applied for tapping mode operation. It is likely that in the near future, commercial AFM instruments will operate in all oscillatory modes and the development of fast, broadband controllers will lead to new multi frequency techniques.


Click on a product type below to order online.

noncontact mode

UHV conditions
HQ:NSC probes for noncontact mode

High-resolution in UHV
Hi'Res-C probes for noncontact mode
TOLL-FREE NUMBER: + 1-866-SPMTIPS (776-8477)
PHONE NUMBER: +49 (0) 6441 2003561
MikroMasch® is a trademark of Nanoworld AG operated by ISB Ltd.