Robust
|
 |
 |
 |
Options |
|
When the sample is relatively robust, i.e. hard and strongly adhering,
this allows imaging without tip-induced deformation in a broad range
of tip-sample interaction force. One may choose using both softer
cantilevers to obtain true surface topography and stiff cantilevers
to ensure high topographic and phase contrast.
|
 |
 |
 |
 |
 |
 |
 |
 |
 |
 |
 |
 |
SEBS60 block-copolymer sample with a
cryomicrotomed surface was imaged
by Tapping Mode AFM under light (top) and hard (bottom) tapping conditions.
The light tapping provide true topography (left), while hard tapping
allows better material contrast in phase images (right). Image courtesy
of Dr. Sergei Magonov (Digital Instruments / Veeco Metrology Group). |
 |
|
The harder the surface, the wider the choice of AFM techniques
can be used for it investigation. Contact mode technique is still
the speediest and roughest among the others.
When in Tapping mode, "hard" tapping conditions are usually
used to obtain good material sensitive contrast in phase imaging,
while "light" tapping keep maximum of topography data
(see figure above).
"Light" tapping conditions imply lower free amplitudes
and higher setpoint values.
|
Contact mode
|
|
Long cantilevers having relatively low spring constant below are
usually used in Contact mode. The choice between SCD or LS tip depends
on resolution needed.
When in liquid, one can use cantilevers without backside coating
or probes with overall chemically inert Cr-Au coating having tip
radius about 50 nm.
|
Tapping mode
|
|
The tip-sample interaction force depends on the adhesive properties
of material, the sharpness of the probe tip and the scanning mode.
If the experimental conditions permit, one should avoid using stiff
cantilevers that may cause severe deformation and even damaging
of the sample. Softer cantilevers usually show better and more reproducible
results. However, in some experiments, e.g. very hard and/or sticky
samples, stiff cantilevers are required. For unknown physical properties
of the domains it is recommended using a probe of the 14 series
that features an intermediate spring constant of ~5 N/m.
Sharper tips provide higher lateral resolution. This advantage
should be balanced against the scanning speed.
For imaging in liquids, the choice of the probe is usually limited
to those having resonant frequency of 50-70 kHz in air. It is generally
recommended to use uncoated cantilevers in liquids or probes with
overall chemically stable Cr-Au coating having tip radius about
50 nm.
|
Noncontact mode
|
|
Cantilevers having high spring constant of 20..100 N/m and highest
possible resonant frequency are usually used in noncontact mode.
|
|
|
|
|




|
MIKROMASCH
TOLL-FREE NUMBERS: USA: 1-866-SPMTIPS (776-8477)/ EU: + 8000-SPMTIPS (776-8477)
E-MAIL: INFO@MIKROMASCH.COM
|