Guide to practical AFM

Options

Besides measurements of some properties, the experiment in AFM may consist in the research of the evolution of the sample in time, or investigation of the impact of a temperature, mechanical forces or chemicals. In many cases researches of the magnetic and electric properties need to be conducted in external electric and magnetic fields.

Topographic tapping mode AFM images (13*13 µm2) show the morphological evolution of a single crystal of PDPS as a function of temperature. Concurrently measured electron-diffraction patterns demonstrate the transformation of the pseudotetragonal R phase of PDPS into the hexagonal columnar mesophase. (* R.I Gearba, et. all., Macromolecules 39, 978-987 (2006)) Images are courtesy of Dr. D. Ivanov.

The experiment in AFM begins with sample preparation, a probe selection, a setting up of a microscope. One should be aware about imaging artifacts that appear as a sharp change of contrast or appearance on identical shapes in various parts of the image. These artifacts can be caused by change of the tip condition and shape due to picking an occasional contamination particle or breaking a tip apex that happens when the tip hit a big structures whose height is larger than a vertical range of the scanner employed in the instruments. Therefore, one should be careful with storage of the samples to avoid various contaminations, and use of the scanners with large range of vertical displacements is preferable, especially, for examination of rough surfaces.

The precise positioning of the probe, which is controlled with an optical viewing of the probe and the sample, may be required for imaging the same area using different probes, or when the probe needs to be in a small sample area (such as single carbon fiber ~ 10 micron in width). In this case it is important to know a position of the tip with respect to the cantilever free end. GP probes are preferable for this reason as the tip is positioned at the end of the cantilever, while in other probes it is offset about 10 micron from the end.

The probe engagement is the important procedure to prevent the tip damage. An example of the careful engagement procedure is described in high-resolution imaging page.

Of course, in quantitative AFM studies it is important to have the microscope scanners properly calibrated. We suggest using calibration gratings TGX01 and step-structures of the TGZ series for proper adjustments of X, Y and Z sensitivity of the scanners.

Depending on a particular task, imaging routines have their specifics. A few of them are described below.
 

Topograpgy imaging

Visualization of surface topography is the main function of AFM. The success of the experiment depends not only on a correctly chosen probe but to a large extent on optimal imaging routine. A minimal tip-sample force, a properly adjusted feedback and an appropriate scanning rate are a combination of important parameters for most precise topographic imaging.

At the end of the topography studies of the particular area, it might be useful to make a final scan on a large area that includes the location where the scans were made. In some cases you might find that imaging of the smaller area left a "window" seen in the large scan. This will be a strong indication that tip-force should be made smaller. How to control the tip-sample force in tapping mode is described in some detais on the topography imaging page.
 

Exploring Materials Morphology and Structural Order

Besides high-resolution topographic information, AFM imaging of ordered structures and heterogeneous materials substantially provides visualization of surface morphology and nanostructures thus complementing other microscopic and diffraction techniques.

In many cases, visualization of ordered structures and compositional imaging of individual components in complex materials are achieved by imaging at elevated forces. This happens in visualization of lamellar order of alkanes and semicrystalline polymers, microphase separation patterns of block copolymers, morphology of elastomer blends, etc. It is not rare that top surfaces of multicomponent polymer materials are enriched of low-molecular weight material and ordered lamellar structures are hidden underneath. In this case, scanning at low tip-force will reveals only stucture of the top layer formed of amorphous polymer and imaging at elevated forces will lead to images showing sub-surface ordered structures. This is the case when imaging at elevated forces might provide images with well-resolved structures contrary to imaging at low forces. For sample with rubbery top layer, which can be penetrated by an AFM probe at elevated forces, one would like to know at what sub-surface depth a particular image was obtained. This information can be gained from analysis of force-curves.

You can find more details on the materials composition mapping page.
 

Examination of local mechanical properties

AFM-nanoindentation experiments can be conducted as follows. First, a sample area where one would like to make indents should be imaged and preferably in the tapping mode to avoid a tip-induced sample modification. When a homogeneous material is studied this allows checking its topography and avoiding occasional surface imperfections that might influence indents. In case of heterogeneous samples a choice of area is important when mechanical properties at interfaces between different components are examined.

While making indents the force curve are collected and later used for extracting force versus penetration curves. Usually, the indents and corresponding force curves are collected in a large number of locations and various levels of the tip-force. The tip indents should be made in different locations, which should be spaced far from each other to exclude interference of the indents. For homogeneous material, this procedure provides the statistically-sound force curve data leading to reliable estimates of elastic modulus. In the examination of the heterogeneous materials, one faces more challenges. First of all, they related to compromises between spacing of indent and strong wish to examine the interfaces between different materials. Second, heterogeneities in bulk demand a more careful assignment of the force curves obtained at different locations. In many a thickness of top material is unknown and the force data might be influenced by another material situated underneath.

After the indents are made, their shapes are visualized with AFM either using the same probe that was employed for indenting or using new and sharper probe. The observation of the indented locations provides important information about the nature of the indent. In case of elastic deformation the material will practically recover and the pits will be barely seen. A partial recovery of the indent takes places for non-elastically deformed materials and a presence of pile-ups at the indent pit indicates a plastic flow developed during indenting. This information should be considered for selection an appropriate model for extracting of quantitative mechanical data from force curves.

In addition to the choice of theoretical model, the quantitative analysis of the force curve requires also knowledge about the prove stiffness, tip geometry and optical sensitivity of the instrument for the particular probe. The probe stiffness is usually measured from probe thermal excitation, cantilevers with pre-measured spring constant are also available (-F Series). The geometry of the tip can be deduced from TEM or SEM micrographs either provided by the probe manufacturer or measured by the AFM user. This is much easy to do for the probe with round-shaped tip apex. The optical sensitivity is usually measured force curves obtained on a rigid substrate (e.g. sapphire) after the indenting experiments. This procedure should be made with an extreme caution to avoid tip damage. Again, in this procedure it will be much easy to handle the probe with rounded tip apex.

The reason why compositional imaging and indenting experiments in AFM are often made on polymer materials is related to the fact that in general stiffness of commercial probes and stiffness of AFM probes are the same range. There are two additional consequences. First of all, due to wide spread of stiffness of polymer materials there should also a wide selection of AFM probes with different stiffness in order to choose the most appropriate probe for compositional imaging of a particular polymer multicomponent material. Second, for nanomechanical examination of biological objects and, particularly, cells one should use the AFM probes which operate at much smaller loads. Due to definite limitations in reducing stiffness of the cantilever one might concentrate on the use of round-shaped beads, which can be attached to the cantilever instead of tips. The beads of micron size can be glued to tip-less probes, like NSC12 and CSC12 cantilevers.

You can find more details on the mechanical properties page.

Studies at Different Temperatures

AFM imaging of polymer samples at different temperatures is used for visualization of phase transitions (melting, crystallization, etc) and direct monitoring of structural changes following thermal annealing. Practically, such imaging is performed after the sample temperature has changed and stabilized. To prevent moisture condensation dry environment of the sample is needed at sub-0C temperatures. This is achieved by purging the heating cell with dry gas. At high temperature purging with inert gas might be need to avoid sample oxidation. In the imaging at elevated temperatures the probe is heated through direct tip-sample contact in the contact mode. In the tapping mode, heating of the probe is made with an additional heater and this prevents a condensation of moisture on the probe that is colder than the sample and condensation of volatile components that might exude from a sample. The heating of the probe make difficult to use the cantilevers with reflected Al coating due to the cantilever bending (bimetallic effect). Additionally, during scanning at high temperature one should be aware about possible increase of sample adhesion. This might restrict a use of soft Si probes. Therefore, for temperature measurements we recommend non-coated Si probes with stiffness ~ 5 N/m and 40 N/m.

Imaging in Different Environments

AFM imaging in gas or vapor atmosphere is usually performed with the same probes that studies at ambient conditions. The situation is different in imaging under liquid, especially, when studies are performed with oscillatory or tapping modes. Due to the fact that driving of the probe in liquid is usually proceeds at the clean resonant frequency the tip-force is not well-controlled. Therefore, in studies under liquid one is limited to soft Si probes with stiffness below 5 nm. Therefore, our probes in the stiffness range from 0.3 N/m to 5 N/m can be used in oscillatory modes in measurements under liquid.

Further reading

Topography imaging

Material composition map

Mechanical properties

Electric properties

Magnetic properties

Contact mode


Ambient or liquid conditions
GP probes for contact mode
DP17/GP/AlBS

Noncontact mode


UHV conditions
GP probes for noncontact mode
DP15/GP/AlBS

Tapping mode


Robust samples in air
GP probes with high spring constant
DP15/GP/AlBS

Soft, weakly-adhering,
or fragile, samples in air

GP probes with medium spring constant
DP14/GP/AlBS

Near-liquid samples in air
GP probes with medium spring constant
DP09/GP/AlBS

High-resolution in air and
imaging of corrugated surfaces

Hi'RES probes with medium spring constant
DP14/Hi'RES-W/AlBS

Low-wear imaging
Hi'RES probes with medium spring constant
DP14/Hi'RES-W/AlBS

Tip shape and radius
characterization

Porous Aluminum structure
PA01

Spectroscopy techniques


Quantitative force measurements
Probes with pre-measured spring constant
-F Series

Indentation and studies of
intermolecular forces

Tipless cantilevers for attaching beads
12 Series

 
 

 

 

 

 
CONTACTS
MIKROMASCH

TOLL-FREE NUMBERS: USA: 1-866-SPMTIPS (776-8477)/ EU: + 8000-SPMTIPS (776-8477)
E-MAIL: INFO@MIKROMASCH.COM