
Characterization and calibration of MFM tip.
Quantitative measurements in Magnetic Force Microscopy.
Theory basics
Since the majority of all Scanning Probe Microscopy techniques is based on
the force or force derivatives detection, the starting relation is:
where Etip-sample is an energy of tip-sample interaction.
Particularly for Magnetic Force Microsocpy Etip-sample
can be expressed in terms of a convolution of a tip magnetization Mtip
and a stray field from the sample Hsample or, vise
versa, as a convolution of the tip stray field Htip and
the sample magnetization Msample:
where integration is performed over the whole magnetic volume of the tip (first
relation) or the sample (the second one). In case of no perturbation between the
tip and the sample both approaches should give equal results.
Therefore, generally speaking, having measured forces
(or their derivatives) acting between a tip and a sample, it is theoretically
possible to restore tip magnetization distribution Mtip if stray
field from the sample Hsample is known in detail. And then,
using so calibrated tip with known Mtip one can extract stray
field distribution from unknown sample using MFM scanning. Vise versa, upon some
model assumptions on distribution of the stray field from MFM tip Htip
and having assessed this value numerically, one can restore magnetisation distribution
Msample within unknown sample. At the same time, care should
be taken in interpreting a stray fields out of the media emanating them because
exactly the same outer distribution of magnetic field can be generated with different
bulk conditions [ 1174].
In most cases, and for MFM in particular, tip-sample interaction is substantially
one-dimensional and consideration of only z-component of force derivative F'z
gives satisfactory numerical estimations:
By now phase shift detection scheme is most widespread
for force gradient mapping in MFM because phase shift is proportional to the force
derivative F':
Besides one-dimensional approximation, a number of another
simplifications are made due to complexity of the problem of magnetic interactions.
First, one can only guess on the real distribution of
magnetisation Mtip in the MFM tip. Therefore,
some simplified models has been proposed to replace real tip with an appropriate
approximation.
There are two most widespread approaches developed to
date: point probe model [ 2649] and extended (magnetic) charge model including
a number of their implementations.
Point probe model assumes that magnetic tip can be replaced
with a single magnetic pole of monopole q or dipole m nature, both
being constant scalar and vector respectively (Fig. 1a).
 |
| Fig. 1. The most widespread models of MFM tip. a) Point probe
model. MFM tip is approximated by a single point pole of monopole q
or dipole m nature. b) Extended charge model. One of the model
implementation is shown. MFM tip is approximated by pyramid with different
magnetizaton vectors on different facets. Active imaging volume is depicted
in white. See also Fig. 3. |
Expression for monopole approximation can be written as:
and for dipole approximation as:
where Meff is effective uniform tip magnetization and
Veff is effective imaging volume of the tip.
Combining both contributions the resulting force in one-dimensional case can be expressed as:
and force derivative as
As a matter of fact, point pole approximation is rather
rough due to above simplifications, and fails to explain entire stray field from
the tip [ 2604]. The advantage is that the field calculation is simple and application
to the theory of MFM is straightforward. This approach is continuously being applied
by many reserchers.
Extended charge model looks much more realistic assuming
that magnetic charges in the tip volume are nonpoint like (Fig. 1b). Unfortunately,
implementations of this model are more complicated. Working in Fourier domain
helps to simplify calculations in this case.
Both models become inaccurate if the samples with characteristic
dimensions of magnetic features different from those used for calibration are
investigated. By now neither model is able to guarantee "absolute" calibration
of the given tip. Thus, recalibration for various samples is necessary.
Limitations of current magnetic force imaging techniques
give rise to refinement of existent theoretical backgrounds and development of
novel experimental approaches such as that described by Schaffer et al. [ 2597].
The process to find out reliable and universal quantification in MFM steadily
goes on.
One should also take into consideration, that measured
MFM signal is a convolution of so-called instrumental response function and pure
MFM signal. So, to extract pure MFM signal specific deconvolution procedures should
be performed. This effect is as much as smaller are the magnetic features on the
sample surface.
Experimental techniques
A number of approaches have been proposed to acquire
both qualitative and quantitative description of magnetic tips in order to make
MFM imaging quantitative. These approaches can be divided into two main groups:
techniques based on MFM scanning over current carrying structures formed on
a calibration sample surface and the others which use another principles involving
MFM for successive correction of initial approximations or do not use MFM at
all.
Let us begin from the second group. This group comprises
Resonant Torque Magnitometry (RTM) [2633, 2621], Lorentz electron tomography
[2603], electron holography [2664] and some other methods.
These methods help understanding peculiarities of magnetic
field distribution in the nearest vicinity of the magnetic tip and are to be
considered as independent ways to develop reliable and simple MFM-like method
of tip characterization/calibration.
Resonant Torque Magnitometry allows a hysteresis loop
to be obtained by applying a local field to the tip, either along the tip axis
or normally to it, and measuring torque response of the tip [2621]. A sensitivity
of better than 10-12 emu (10-15 A·m2)
has been reported by Heiydon et. al [2633]. Measurements were performed in-situ
that is of great importance especially when images are made in the presence
of bias fields. At the same time this technique is rather exotic and cannot
be implemented in common laboratories. RTM also impose some limitations to the
measuring conditions.
Lorentz electron tomography used by McVitie et al. [2603]
registers deflection of accelerated electron beam influenced by magnetic field
in the nearest vicinity of MFM tip (Fig. 2).
 |
| Fig. 2. Schematic of differential phase contrast imaging technique
of Lorentz microscopy applied to a MFM tip (McVitie et al. [2603]). This
imaging mode is implemented on a scanning transmission electron microscope. |
This technique gives the real, but integral field distribution
near the tip apex not distorted by any external fields. The data obtained easily
reveal anisotropy of the field around the axis of the tip due to its nonideal
piramidal form. Small tilt over the measured surface as well as some unwanted
effects ([ 2663]) also contribute to the final result. Then, specific tomographic
reconstruction is implemented to restore the real field distribution. The authors
compared data obtained by electron tomography with the results of calculations
according to most common approximations of MFM tip magnetic state such as point
pole and extended charge models. It was found that the field from MFM tip has
contributions which are point like and nonpoint like. While the central part of
the field can be well represented by a point or localized charge, the modeling
of extended tails of the field implies nonlocal magnetic charge distribution.
The profile of magnetic field using extended charge model (Fig. 3) can be ideally
fitted by means of fitting parameters. There are, though, some difficulties in
interpretation of thus obtained results.
 |
| Fig. 3. Representation of MFM tip by magnetic layer of thickness
t with predefined magnetization distribution according to extended
charge model used by McVitie et al. [2603] for Fourier space calculations. |
Another technique involving multiple MFM sessions
and Fourier transforms is developed by Hug, van Schendel et al. [ 2632, 2604].
It is a sophisticated iterative procedure for MFM tip calibration using transfer-function
approach. The sample under scanning is specialty sandwich-like Cu/Ni/Cu structure
with strong perpendicular magnetization. Authors came to conclusion that the
point monopole and dipole models of the MFM tip are not suitable to describe
the tip calibrations found in the experiment. And in spite of the fact that
a reasonable agreement between the simulation and the measurement is obtained
with a monopole model with additional tip-sample distance, it is not recommended
to use a point pole model for the evaluation of MFM data acquired on a sample
with domains with a different size than those of the sample used for calibration.
Even using extended charge model one should recalibrate the tip to investigate
samples with magnetic feature sizes different from used for calibration. The
authors found point pole probe approach inadequate when interpreting submicron
magnetic features which are comparable to the tip size, though this approach
may be still applicable for the structures with relatively large features and
large lift heights. The maximum magnetic flux density at the tip end was estimated
about 0,01 Tesla for about 10 nm Fe layer covered with about 15 nm Au coatings.
The sensitivity is found to be in the microtesla range for stray field variations
on the order of 100 nm.
By common practice, determination of a sample stray field
using calibrated MFM tip of given magnetization Mtip is the
most often task investigators have to encounter. This task is successfully solved
using MFM tips calibrated by means of the first group techniques. It is obvious,
that only current carrying metal stripes are able to produce well-controllable
and regular magnetic field with predefined distribution Hsample
to restore tip magnetization Mtip. Due to micron and submicron
scales these stripes are manufactured by electron beam lithography and may differ
in geometry. The following types of strip lines are reported so far: one straight
strip line [ 2582, 2619];
two or more straight strip lines [ 2599, 2600,
2631, 2634,
2635, 2639],
current rings [ 2612, 2636,
2637].
 |
| Fig. 4. Typical current strip line configurations for generating
of well-defined magnetic field. a) Two parallel strip lines with regions
of various distance between lines (Goddenhenrich et al. [2631]). b) Current
ring (Kong and Chou [2636, 2637], Lohau et al. [2612]). |
The first attempt to compare calculated and measured magnetic
forces acting between a tip and a sample using current carrying strip lines was
undertaken by Goddenhenrich et al. [2631] in early 90-s (Fig. 4a)). The authors
have got satisfactory qualitative agreement of calculated and measured curves.
To get semiquantitative results convolution effects due to finite tip size should
be also taken into consideration.
The distribution of magnetic field gradient for a couple
of straight lines is shown in Fig. 5. Simple and comprehensible animations help
to understand how magnetic field and its gradient are distributed over the current
carrying lines.
 |
Fig. 5. Distribution of horizontal Hx and vertical Hy components
of magnetic field gradient over two current carrying strip lines of opposite
polarity upon height Z in microns. The lines go in parallel to each other
and their cross sections are only shown. Both components are plotted separately
for convenience. Click on the picture to download the full animaiton (80
Kb in a self-extracting archive).
One can see that distribution of Hx component
is symmetrical relative to the strip line profile, and that of Hz is antisymmetrical.
After viewing the full animation it is obvious that the distribution of
Hx component changes drastically upon vertical distance Z as opposed to
Hy which changes in amplitude only. |
Yongsunthon et al. [2599] consider MFM as a useful method
for studying electromigration in metals which is of vital importance due to continuous
shrinkage of integrated circuit dimensions. Proposed calibration structure comprises
three metal strip lines (Fig. 6), of which the left is null-potential grounded
line connected to the tip, the central line includes small slit in the central
part thus modeling defect, and the right line is reference one of well defined
geometry.
 |
| Fig. 6. Current strip line configuration used for in situ calibration
of MFM tip by Yongsunthon et al. [2599]. |
Null-potential line guarantees the absence of electrostatic
component among various tip-sample interactions. It serves to indicate that no
phase shift is present when the tip is scanning above this area. The reference
structure is used to normalize the signal magnitude from the test structure. Also
it is intended to determine iteratively so-called instrumental response function
which enables deconvolution of measured MFM signal in order to extract pure MFM
response. Using dipole moment approximation in accordance with point probe model,
relative quantification of the MFM signal amplitude to within 10% can be achieved
and absolute current variations can be detected to at least 10%.
Successful application of MFM imaging to map quantitatively
current densities around various defects in metal conductors is reported in the
latest works of Yongsunthon et al. [2639] and Rose et al. [2600].
Results presented lately by Kebe and Carl [2634] reflect
progress made for more than a decade in MFM tip calibration using current carrying
structures. Beginning from current rings in early works, the authors consider
various parallel wire configurations in this ample paper. Analysis is again conducted
using point probe model for approximation of MFM tip. Having studied many configurations
of current carrying lines authors related point probe parameters to the dimensions
of parallel wires and to characteristic decay length of the z-component of the
magnetic field produced with them. This allowed to determine the effective volume
Veff of the real magnetic tip relevant in MFM imaging. The authors
proposed experimental technique which allows quantitative measurement of the magnetization
of nanoscale ferromagnetic elements with an in-plane magnetization by calibrated
MFM tip.
Kong and Chou have pioneered in using current rings for
MFM tip calibration purposes [2636, 2637]. Using monopole and dipole approximations
within a point pole model the authors found that at different lift heights both
moments contribute unequally, which gave rise to the final conclusion that only
the dipole moment is relevant for imaging.
Lohau et al. [2612] thoroughly studied applicability of
point pole model to the quantitative determination of magnetic state of the tip
using current rings of various radii. It was found that MFM images can only be
analyzed unambiguously within the point pole approximation when using either the
dipole or the monopole contribution taking into account heights of the point pole
above the sample surface which are different in each case. Both point probe approaches
was found to give identical results independent of the lift height used and it
is therefore not necessary to consider a mixture of both contributions.
A number of papers describe changes in magnetization of
MFM tips imposed by external magnetic fields. Use of current carrying structures
becomes here the only way to determine tip magnetic state because any other magnetic
sample is critically affected under strong fields applied, whereas the stray field
from such current carrying structures is kept unchanged. Knowing magnetization
reversal of the tip one can study magnetization reversal of magnetic samples.
Typical hysteresis loops for tips of various shape and
covered with various coatings are presented in Fig. 7.
 |
| Fig. 7. Typical hysteresis loops obtained for various metal coatings
of MFM tips. High coercivity coating has a broad hysteresis loop (solid
red line). Low coercivity coating has a narrow loop (dashed red line).
Blue line reflects behavoir of coating deposited on narrow and long tip.
This coating switches as a single domain particle under varying external
field and features almost square hysteresis loop. Hc, Mr
and Ms stand for coercivity, remanent magnetization and saturation
magnetization respectively. |
In general, magnetic coatings can be divided into two
categories: made of high coercivity (CoCr, CoCrPt etc.) and low coercivity (FeNi,
FeCoNi etc.) materials. There were also reports about superparamagnetic [2730,
2567], paramagnetic [2705],
antiferromagnetic [2711] or multilayer [2691]
MFM tips. The proper choice depends upon the purposes of a given study. It is
found experimentally that typical coercivity of CoCr tips is between 300-400
Oe, and for low-coercivity tips Hc amounts to 1-5 Oe or less.
One should remember, that high coercivity tips may influence
the sample magnetic structure and vice versa. Such perturbation leads to displacement
of domain walls [2615, 2748,
2749], irreproducibility of successively taken
images [1111, 1122],
and another unwanted effects. As for imaging in applied external magnetic fields,
it is desirable to use a tip either having coercivity far beyond that of the
sample (so that the applied field does not affect the tip), or far below it
(so that the tip moment is always aligned with applied field) [2582]. Tapping
the sample with the MFM tip during topography acquisition in interleave or "liftmode"
may significantly perturb magnetization distribution in the sample or in the
tip. Therefore, if sample topography is sufficiently smooth, noncontact imaging
at fixed height is preferable instead of two-pass technique such as liftmode.
The more aspect ratio of the tip, the more uniform and
predictable is magnetization distribution in the volume of magnetic coating.
The utmost case is narrow needle shaped tip proved to be monodomain, so hysteretic
loop of such a tip is of almost square shape (so-called coefficient of squareness
is close to 1.00).
Babcock et al. [2582] systematically studied the magnetic
states of a set of MFM probes as a function of uniform external magnetic field
using single current strip line. In order to avoid influence of electrostatic
forces entire structure was covered by thin insulating and metal coatings, the
latter being grounded during MFM measurements. It was shown that there are two
different response characteristics for lateral and vertical components of remanent
magnetization in the phase images. Vertical component gives antisymmetric response
and the lateral - symmetric one. Such an effect is easily seen from the field
gradient profiles, plotted in Fig. 5. The authors show that after magnetizing
even in a strong lateral fields remanent state is apparently far from horizontal
orientation but tilted 30-40° relative to the plane of the sample.
 |
|
Fig. 8. Responses of the MFM tip scanning the single strip line
(shown above) under various orientations of external field. (after Babcock
at al. [2582])
|
Carl et al. [ 2569] describes
an experimental technique to determine magnetization reversal and coercivity of
MFM tips using current carrying metal ring. Measurements were conducted in the
central part of substantially large (Ø=2400 nm) ring. Such a big size guarantees
that mainly vertical H z component is present in the central part of
the ring and there are no electrostatic forces present in this area, so shielding
metal coating was not exploited. Coercivity Hc, remanent magnetization
Mr, and saturation magnetization Ms of the
tips from corresponding hysteresis loops were obtained changing external field
from Hz to -Hz.
It was found that properties of theoretically identical tips from various wafers
differ widely, sometimes by factor of 2. This discrepancy can only be attributed
to unequal manufacturing conditions rather than to other reasons. Typical coercivity
for CoCr coated tips were found to be within 270-360 Oe and remanent magnetization
Mr=869-1336 emu/cc. For comparison the same characteristics
were gathered by Superconducting Quantum Interference Magnitometry (SQUID Magnetometry).
Values obtained by SQUID measurements differed from that of MFM widely. For
example, Hc,SQUID/Hc,MFM
relation lie within 0,9??¦2,2 range. It is only MFM that allow estimation of the
volume of the tip relevant for image formation. Therefore, the choice in favour
of MFM rather than SQUID magnetometry is preferable to assess most common magnetic
properties of the probing tip.
Using so calibrated tip Lohau, Carl et al. [2602]
determined quantitatively the hysteresis loop of a single magnetic dot of 230
nm in diameter taking into account the correction for the tip reversal properties.
The technique described in the paper allows one to measure precisely the coercivity
of a given sample Hc,sample even if it is comparable with
the coercivity of the tip itself. Furthermore, the shape of the hysteresis loop
of small magnetic elements may be obtained with sufficient accuracy in order
to be able to differentiate, e.g., between different reversal modes of magnetic
elements.
In conclusion, some general regularities will be given
concerning resolution and sensitivity in MFM measurements.
Probe sensitivity over the current strip line can be
estimated in terms of minimal magnetic moment within point dipole approximation:

where f o - resonant frequency shift. For example, in case of most commonly
used CoCr coatings mz of ~ 6?·10-12 emu has
been reported [2582].
MFM resolution and magnitude of the contrast observed
in MFM depend upon the geometry and size of the magnetic material which interacts
with the sample stray field. Such working part is also called "active volume"
of the MFM tip (depicted in white in Fig 1b). Both resolution and contrast for
"magnetically hard" MFM probes increase as the final portion of the
tip goes long and narrow having flat end [2645].
Such tips are manufactured using Electron Beam Deposition [2643,
2728] or Focused Ion Beam [2647,
2691] techniques. At the same time one should remember, that such tips are easily
worn during tapping while acquiring topography by means of widely used two-pass
technique. So, noncontact mode is preferable if no great topography variations
are present on the sample surface.
Horizontal and vertical resolution of MFM imaging differs
greatly. Commercial tips are intended mainly for sensing of vertical field component
whereas in-plane components are often out of their capabilities. To make tip
sensitive to lateral forces, one should manage to make the very end of the tip
magnetized horizontally. Simple applying of external magnetic field in horizontal
direction will give a little success. The remedy could be making nanoscopic
pinhole in the very apex of the magneitc tip as offered by Folks et al. [2610].
This lead to occurrence of very small region with sufficiently strong horizontal
magnetic moment. Another three- step way to extract information about distribution
of lateral magnetic moments from MFM scans is described by Zhao et al. [2648].
Abelmann et al. [1108]
reported results of a great study devoted to comparing the resolution of MFM
from various manufacturers using special reference samples consisting of CoNi/Pt
magneto-optic multilayers with different thicknesses. It was found that resolution
does not vary considerably between the different instruments and lies within
30-100 nm range. Theoretical analysis shows that maximum resolution is governed
by the distance between magnetic charges in the sample and in the tip which
is quite different from the physical tip-sample separation. The lower the separation
the higher resolution. Though, reduction of physical tip-sample separation leads
to a plenty of smaller details or even artefacts of nonmagnetic origin. Therefore,
reduction of magnetic rather than physical distance between interacting charges
becomes a matter of art.
Dependence of MFM data on the tip magnetization orientation
has been studied by Litvinov and Khizroev [2709],
who manufactured the finest cylindrical MFM tip of 50 nm in diameter and 10
nm in height which is one of the best ever reported physical implementation
of the "dipole moment" MFM probe. By controlling the preferred orientation
of the magnetization, it is possible to define directional sensitivity. The
preferred orientation is controlled either by choosing a proper magnetic material
with preferred crystalline anisotropy or by applying a sufficiently strong external
magnetic field.
Complete list of references to the articles devoted
to MFM-related problems is placed in the Reference
Collections section of our Library.
| ID |
Reference list (newly come references are marked red) |
| 2649 |
The point dipole approximation in magnetic force microscopy
U. Hartmann
Phys. Lett. A, 137 (1989) 9, 475-478 |
| 2631 |
Probe calibration in magnetic force microscopy
T. Goddenhenrich, H. Lemke, M. Muck, U. Hartmann, and C. Heiden
Appl. Phys. Lett., 57 (1990) 24, 2612-2614 |
| 1185 |
The influence of experimental parameters on contrast formation in magnetic
force microscopy
F. Kaisinger, H. Starke, G. Persch, U. Hartmann, F. Krause
Thin Solid Films, 264 (1995) 2, 141-147 |
| 1185 |
The influence of experimental parameters on contrast formation in magnetic
force microscopy
F. Kaisinger, H. Starke, G. Persch, U. Hartmann, F. Krause
Thin Solid Films, 264 (1995) 2, 141-147 |
| 1120 |
High-resolution magnetic imaging based on scanning probe techniques
U. Hartmann
Journal of Magnetism and Magnetic Materials, 157-158 (1996) 545-549 |
| 2664 |
Quantitative magnetometry using electron holography: field profiles
near magnetic force microscope tips
D.G. Streblechenko, M.R. Scheinfein, M. Mankos, K. Babcock
IEEE Trans. Magn., 32 (1996) 5, 4124-4129 |
| 1135 |
Magnetic force microscopy images of high-coercivity permanent magnets
L. Folks, R. Street, R.C. Woodward, K. Babcock
Journal of Magnetism and Magnetic Materials, 159 (1996) 1-2, 109-118 |
| 2582 |
Field-dependence of microscopic probes in magnetic force microscopy
K. L. Babcock, V. B. Elings, J. Shi, D. D. Awschalom, and M. Dugas
Appl. Phys. Lett., 69 (1996) 705-707 |
| 1122 |
Interactions between soft magnetic samples and MFM tips
S.L. Tomlinson, A.N. Farley, S.R. Hoon, M.S. Valera
Journal of Magnetism and Magnetic Materials, 157-158 (1996) 557-558 |
| 2730 |
Superparamagnetic magnetic force microscopy tips
P. F. Hopkins, J. Moreland, S. S. Malhotra, S. H. Liou
J. Appl. Phys., 79 (1996) 8, 6448-6450 |
| 2650 |
Quantitative Magnetic Field Measurements With The Magnetic Force Microscope
Roger Proksch, George D. Skidmore, E. Dan Dahlberg, Sheryl Foss, J. J. Schmidt,
and Chris Merton, Brian Walsh, Matt Dugas
Appl. Phys. Lett., 69 (1996) 17, 2599-2601 |
| 2728 |
Electron beam fabrication and characterization of high-resolution magnetic
force microscopy tips
M. Ruhrig, S. Porthun, and J. C. Lodder, S. McVitie, L. J. Heyderman, A.
B. Johnston, and J. N. Chapman
J. Appl. Phys., 79 (1996) 6, 2913-2919 |
| 2745 |
Investigation of the domain contrast in magnetic force microscopy
L. Belliard, A. Thiaville, S. Lemerle, A. Lagrange, J. Ferre, and J.
Miltat
J. Appl. Phys., 81 (1997) 8, 3849-3851 |
| 2643 |
Improved spatial resolution in magnetic force microscopy
G. D. Skidmore, E. D. Dahlberg
Appl. Phys. Lett., 71 (1997) 3293-3295 |
| 2733 |
A microfabricated tip for simultaneous acquisition of sample topography
and high-frequency magnetic field
V. Agrawal, P. Neuzil, and D. W. van der Weide
Appl. Phys. Lett., 71 (1997) 16, 2343-2345 |
| 2748 |
Magnetic dissipation force microscopy
P. Grutter, Y. Liu, P. LeBlanc, and U. Durig
Appl. Phys. Lett. , 71 (1997) 2, 279-281 |
| 2636 |
Quantification of magnetic force microscopy using a micronscale current
ring
Linshu Kong and Stephen Y. Chou
Appl. Phys. Lett., 70 (1997) 15, 2043-2045 |
| 2637 |
Study of magnetic properties of magnetic force microscopy probes using
micronscale current rings
Linshu Kong and Stephen Y. Chou
J. Appl. Phys., 81 (1997) 8, 5026-5028 |
| 2638 |
Micromagnetic model for magnetic force microscopy tips
S. L. Tomlinson and A. N. Farley
J. Appl. Phys., 81 (1997) 8, 5029-5031 |
| 2633 |
Resonant torque magnetometry: a new in-situ technique for determining
the magnetic properties of thin film MFM tips
G.P. Heydon, A.N. Farley, E. Hoon, O. Valera, S.L. Tomlinson
IEEE Trans. Magn., 33 (1997) 5, 4059-4061 |
| 2568 |
Measurement of the stray field emanating from magnetic force microscope
tips by Hall effect microsensors
A. Thiaville, L. Belliard, D. Majer, E. Zeldov, J. Miltat
J. Appl. Phys., 82 (1997) 7, 3182-3191 |
| 2567 |
High resolution imaging of thin-film recording heads by superparamagnetic
magnetic force microscopy tips
S. H. Liou, S. S. Malhotra, John Moreland, P. F. Hopkins
Appl. Phys. Lett., 70 (1997) 1, 135-137 |
| 873 |
Magnetically refined tips for Scanning Force Microscopy
R. Jumpertz, P. Leinenbach, A.W.A. van der Hart, J. Schelten
Microelectronic Engineering, 35 (1997) 1-4, 325-328 |
| 2632 |
Quantitative magnetic force microscopy on perpendicularly magnetized
samples
Hans J. Hug, B. Stiefel, P. J. A. van Schendel, A. Moser, R. Hofer, S. Martin,
H.-J. Guntherodt, S. Porthun, L. Abelmann, J. C. Lodder, G. Bochi, and R.
C. O'Handley
J. Appl. Phys., 83 (1998) 11, 5609-5620 |
| 1174 |
On the determination of the internal magnetic structure by magnetic
force microscopy
B. Vellekoop, L. Abelmann, S. Porthun, C. Lodder
Journal of Magnetism and Magnetic Materials, 190 (1998) 1-2, 148-151 |
| 1111 |
Development of high coercivity magnetic force microscopy tips
S.H. Liou, Y.D. Yao
Journal of Magnetism and Magnetic Materials, 190 (1998) 1-2, 130-134 |
| 1123 |
Interpretation of low-coercivity tip response in MFM imaging
R. Street, D.L. Bradbury, L. Folks
Journal of Magnetism and Magnetic Materials, 177-181 (1998) 980-981 |
| 1108 |
Comparing the resolution of magnetic force microscopes using the CAMST
reference samples
L. Abelmann, S. Porthun, M. Haast, C. Lodder, A. Moser, M.E. Best, P.J.A.
van Schendel, B. Stiefel, H.J. Hug, G.P. Heydon, A. Farley, S.R. Hoon, Thomas
Pfaffelhuber, R. Proksch, and K. Babcock
Journal of Magnetism and Magnetic Materials, 190 (1998) 1-2, 135-147 |
| 2596 |
Optimization of lateral resolution in magnetic force microscopy
S. Porthun, L. Abelmann, S.J.L. Vellekoop, J.C. Lodder, H.J. Hug
Applied Physics A: Materials Science & Processing, 66 (1998) 7, S1185-S1189 |
| 1111 |
Development of high coercivity magnetic force microscopy
tips
S.H. Liou, Y.D. Yao
Journal of Magnetism and Magnetic Materials, 190 (1998), 1-2, 130-134
|
| 1174 |
On the determination of the internal magnetic structure
by magnetic force microscopy
B. Vellekoop, L. Abelmann, S. Porthun, C. Lodder
Journal of Magnetism and Magnetic Materials, 190 (1998), 1-2, 148-151
|
| 2749 |
Magnetic dissipation force microscopy studies of magnetic materials
Y. Liu and P. Grutter
J. Appl. Phys., 83 (1998) 11, 7333-7338 |
| 2666 |
Comparing the resolution of magnetic force microscopes using the CAMST
reference samples
L. Abelmann, S. Porthun, M. Haast, C. Lodder, A. Moser, M. E. Best,
P.J.A. van Schendel, B. Stiefe, H. J. Hug, G.P. Heydon, A. Farley, S.R.
Hoon, Th. Pfaffelhuber, R. Proksch, K. Babcock
Journal of Magnetism and Magnetic Materials, 190 (1998) 135-147 |
| 2619 |
Quantification of magnetic force microscopy images using combined electrostatic
and magnetostatic imaging
R. D. Gomez, A. O. Pak, A. J. Anderson, E. R. Burke, A. J. Leyendecker,
and I. D. Mayergoyz
J. Appl. Phys., 83 (1998) 11, 6226-6228 |
| 1117 |
Fabrication and characterization of advanced probes for magnetic force
microscopy
U. Hartmann, J. Schelten, P. Leinenbach, U. Memmert
Applied Surface Science, 144-145 (1999) 492-496 |
| 1147 |
Magnetic microscopies: the new additions - The Analysis of Magnetic
Microstructures
R. Proksch, E. Dan Dahlberg
Journal of Magnetism and Magnetic Materials, 200 (1999) 1-3, 720-728 |
| 2611 |
Range of interactions: An experiment in atomic and magnetic force microscopy
W. L. Murphy et al.
Am. J. Phys., 67 (1999) 905 |
| 2612 |
Quantitative determination of effective dipole and monopole moments
of magnetic force microscopy tips
J. Lohau, S. Kirsch, A. Carl, G. Dumpich, and E. F. Wassermann
J. Appl. Phys., 86 (1999) 3410-3417 |
| 1104 |
Calculation of playback signals from MFM images using transfer functions
S.J.L. Vellekoop, J.J. Miles, J.C. Lodder, L. Abelmann, S. Porthun
Journal of Magnetism and Magnetic Materials, 193 (1999) 1-3, 474-478 |
| 1116 |
Extracting media noise characteristics from MFM images
T. Minvielle, S. Nair, P. Arnett
Journal of Magnetism and Magnetic Materials, 193 (1999) 1-3, 479-483 |
| 1176 |
Preparation and characterisation of a new amorphous tip coating for
application in magnetic force microscopy
H.A. Davies, S. McVitie, M.R.J. Gibbs, R.P. Ferrier, W.M. Rainforth, J.
Scott, G.P. Heydon, J.W. Tucker, J.E.L. Bishop
Journal of Magnetism and Magnetic Materials, 205 (1999) 2-3, 131-135 |
| 1175 |
Potentiometric and magnetic force microscopy in multilayers
D.B. Lambrick, M.A. Slade, H. Holloway, M.S. Valera, A.N. Farley, D.J.
Kubinski, S.R. Hoon
Journal of Magnetism and Magnetic Materials, 198-199 (1999) 95-97 |
| 2614 |
Observation of the effects of tip magnetization states on magnetic
force microscopy images
Paul Rice and Stephen E. Russek
J. Appl. Phys., 85 (1999) 8, 5163-5165 |
| 2615 |
Magnetic dissipation microscopy in ambient conditions
Roger Proksch, Ken Babcock, and Jason Cleveland
Appl. Phys. Lett., 74 (1999) 419-421 |
| 1110 |
Description of magnetic force microscopy by three-dimensional tip Green's
function for sample magnetic charges
H. Saito, S. Ishio, J. Chen
Journal of Magnetism and Magnetic Materials, 191 (1999) 1-2, 153-161 |
| 2621 |
Single layer and multilayer tip coatings in magnetic force microscopy
S. M. Casey, D. G. Lord, P. J. Grundy, M. Slade, and D. Lambrick
J. Appl. Phys., 85 (1999) 8, 5166-5168 |
| 2605 |
Quantitative determination of the magnetization and stray field of
a single domain Co/Pt dot with magnetic force microscopy
J. Lohau, S. Kirsch, A. Carl, and E. F. Wassermann
Appl. Phys. Lett., 76 (2000) 3094-1096 |
| 2604 |
A method for the calibration of magnetic force microscopy tips
P. J. A. van Schendel, H. J. Hug, B. Stiefel, S. Martin, and H.-J. Guntherodt
J. Appl. Phys., 88 (2000) 435-445 |
| 1126 |
Investigation of the response of a new amorphous ferromagnetic MFM
tip coating with an established sample and a prototype device
G.P. Heydon, W.M. Rainforth, M.R.J. Gibbs, H.A. Davies, J.E.L. Bishop,
J.W. Tucker, S. Huo, G. Pan, D.J. Mapps, W.W. Clegg
Journal of Magnetism and Magnetic Materials, 214 (2000) 3, 225-233 |
| 2610 |
Perforated tips for high-resolution in-plane magnetic force microscopy
L. Folks, M. E. Best, P. M. Rice, B. D. Terris, D. Weller, and J. N.
Chapman
Appl. Phys. Lett., 76 (2000) 909-911 |
| 1825 |
Carbon-nanotube probe equipped magnetic force microscope
T. Arie, H. Nishijima, S. Akita and Y. Nakayama
J. Vac. Sci. Technol., B18 (2000) 1, 104-106 |
| 1207 |
Low temperature magnetic force microscopy with enhanced sensitivity
based on piezoresistive detection
A. Volodin, K. Temst, C. Van Haesendonck, and Y. Bruynseraede
Rev. Sci. Instrum., 71 (2000) 12, 4468-4473 |
| 642 |
Non-contact atomic force microscopy of an antiferromagnetic NiO(100)
surface using a ferromagnetic tip
H. Hosoi, M. Kimura, K. Hayakawa, K. Sueoka, K. Mukasa
Applied Physics A: Materials Science & Processing, 72 (2001) 7, S23-S26 |
| 1835 |
Quantitative Analysis of the Magnetic Properties of a Carbon Nanotube
Probe in Magnetic Force Microscopy
T. Arie, N. Yoshida, S. Akita and Y. Nakayama
J. Phys. D: Appl. Phys., 34 (2001) L43-L45 |
| 2602 |
Magnetization reversal and coercivity of a single-domain Co/Pt dot
measured with a calibrated magnetic force microscope tip
J. Lohau, A. Carl, S. Kirsch, and E. F. Wassermann
Appl. Phys. Lett., 78 (2001) 14, 2020-2022 |
| 2603 |
Quantitative field measurements from magnetic force microscope tips
and comparison with point and extended charge models
S. McVitie, R. P. Ferrier, J. Scott, G. S. White, and A. Gallagher
J. Appl. Phys., 89 (2001) 7, 3656-3661 |
| 2598 |
Quantitative interpretation of magnetic force microscopy images from
soft patterned elements
J. M. Garcia et al.
Appl. Phys. Lett. 79 (2001) 656 |
| 1148 |
Magnetic phase transitions studied by magnetic force microscopy
Y.-A. Soh, G. Aeppli, N.D. Mathur, M.G. Blamire
Journal of Magnetism and Magnetic Materials, 226 (2001) 857-859 |
| 2714 |
Quantitative interpretation of magnetic force microscopy images from
soft patterned elements
J. M. Garcia, A. Thiaville, J. Miltat, K. J. Kirk, J. N. Chapman, and F.
Alouges
Appl. Phys. Lett., 79 (2001) 5, 656-658 |
| 2734 |
Computer Simulation of Magnetic Force Microscopy Images with a Static
Model of Magnetization Distribution and Dipole-Dipole Interaction
D. V. Ovchinnikov and A. A. Bukharaev
Tech. Phys., 46 (2001) 8, 1014 |
| 2705 |
Micromagnetic structure images taken using platinum coated tips
O. Teschke
Appl. Phys. Lett., 79 (2001) 17, 2773-2775 |
| 2742 |
High resolution eddy current microscopy
M. A. Lantz, S. P. Jarvis, and H. Tokumoto
Appl. Phys. Lett., 78 (2001) 3, 383-385 |
| 2752 |
Magnetic-field measurements of current-carrying devices by force-sensitive
magnetic-force microscopy with potential correction
Tony Alvarez, Sergei V. Kalinin, and Dawn A. Bonnell
Appl. Phys. Lett. , 78 (2001) 7, 1005-1007 |
| 2648 |
Reconstruction of in-plane magnetization distributions from magnetic
force microscope images
T. Zhao, H. Fujiwara, G. J. Mankey, C. Hou, and M. Sun
J. Appl. Phys., 89 (2001) 11, 7230-7232 |
| 2663 |
Electrostatic charging artefacts in Lorentz electron tomography of
MFM tip stray fields
J. Scott, S. McVitie, R.P. Ferrier and A. Gallagher
Journal of Physics D: Applied Physics, 34 (2001) 9, 1326-1332 |
| 2737 |
Current-modulating magnetic force microscope probe
Frank Z. Wang, Na Helian, Warwick W Clegg, James F. C. Windmill, and
David Jenkins
J. Appl. Phys., 89 (2001) 11, 6778-6780 |
| 2599 |
Calibrated magnetic force microscopy measurement of current-carrying
lines
R. Yongsunthon, J. McCoy, and E. D. Williams
J. Vac. Sci. Technol., A19 (2001) 4, 1763-1768 |
| 2569 |
Magnetization reversal and coercivity of magnetic-force microscopy
tips
A. Carl, J. Lohau, S. Kirsch, and E. F. Wassermann
J. Appl. Phys., 89 (2001) 11, 6098-6104 |
| 2635 |
Calibration of magnetic force microscopy using micron size straight
current wires
Congxiao Liu, Kingston Lin, Rich Holmes, Gary J. Mankey, Hideo Fujiwara,
Huaming Jiang, and Hae Seok Cho
J. Appl. Phys., 91 (2002) 10, 8849-8851 |
| 1382 |
Self-sensing piezoresistive cantilever and its magnetic force microscopy
applications
Hiroshi Takahashi, Kazunori Ando and Yoshiharu Shirakawabe
Ultramicroscopy, 91 (2002) 1-4, 63-72 |
| 2691 |
Magnetic force microscopy using focused ion beam sharpened tip with
deposited antiferro-ferromagnetic multiple layers
Zhiyong Liu, You Dan, Qiu Jinjun, and Yihong Wu
J. Appl. Phys., 91 (2002) 10, 8843-8845 |
| 2647 |
High resolution magnetic force microscopy using focused ion beam modified
tips
G. N. Phillips, M. H. Siekman, L. Abelmann, J. C. Lodder
Appl. Phys. Lett., 81 (2002) 5, 865-867 |
| 2731 |
Test of response linearity for magnetic force microscopy data
R. Yongsunthon, E. D. Williams, J. McCoy, R. Pego, A. Stanishevsky,
P. J. Rous
J. Appl. Phys., 92 (2002) 3, 1256-1261 |
| 2722 |
The extraction of features from magnetic force microscopy images using
neural network techniques
H. V. Jones and R. W. Chantrell
J. Appl. Phys., 91 (2002) 10, 8855-8857 |
| 2712 |
Quantitative analysis of transition curvature by magnetic force microscopy
Feng Liu, Shaoping Li, Yan Liu, George Gray, and Allan Schultz
J. Appl. Phys., 91 (2002) 10, 6842-6844 |
| 2709 |
Orientation-sensitive magnetic force microscopy for future probe storage
applications
Dmitri Litvinov and Sakhrat Khizroev
Appl. Phys. Lett., 81 (2002) 10, 1878-1880 |
| 2428 |
Spin-polarized scanning tunneling microscopy with antiferromagnetic
probe tips
A. Kubetzka, M. Bode, O. Pietzsch, R. Wiesendanger
Phys. Rev. Lett., 88 (2002) 5, 57201 |
| 2198 |
Orientation-sensitive magnetic force microscopy for future probe storage
applications
Dmitri Litvinov and Sakhrat Khizroev
Appl. Phys. Lett., 81 (2002) 1878-1880 |
| 2639 |
Mapping electron flow using magnetic force microscopy
R. Yongsunthon, A. Stanishevsky, E. D. Williams, and P. J. Rous
Appl. Phys. Lett., 82 (2003) 19, 3287-3289 |
| 2597 |
Magnetic force gradient mapping
Tilman E. Schaffer, Manfred Radmacher, and Roger Proksch
J. Appl. Phys., 94 (2003) 10, 6525-6532 |
| 2645 |
High-resolution MFM: simulation of tip sharpening
H. Saito, A. G. vanden Bos, L. Abelmann, J. C. Lodder
IEEE Trans Magn, 39 (2003) 5, 3447-3449 |
| 2711 |
Point-dipole response from a magnetic force microscopy tip with a synthetic
antiferromagnetic coating
Yihong Wu, Yatao Shen, Zhiyong Liu, Kebin Li, and Jinjun Qiu
Appl. Phys. Lett., 82 (2003) 11, 1748-1750 |
| 2600 |
Real-space imaging of current distributions at the submicron scale
using magnetic force microscopy: Inversion methodology
P. J. Rous, R. Yongsunthon, A. Stanishevsky, and E. D. Williams
J. Appl. Phys., 95 (2004) 5, 2477-2486 |
| 2634 |
Calibration of magnetic force microscopy tips by using nanoscale current-carrying
parallel wires
Th. Kebe and A. Carl
J. Appl. Phys., 95 (2004) 3, 775-792 |
|